Professional Chamber Furnaces with Brick or Fibre Insulation
 
 LH 15/12 with brick insulation LH 60/12 with scale to measure weight reduction during annealing

LF 120/12 with fibre insulation

LH 120/12S with quartz glass retort

Parallel swinging door for opening when hot

Gas panel

LH 15/12 - LF 120/14

The LH 15/12 - LF 120/14 laboratory furnaces have been trusted for many years as professional chamber furnaces for the laboratory. These furnaces are available with either a robust insulation of light refractory bricks (LH models) or with a combination insulation of refractory bricks in the corners and low heat storage, quickly cooling fibre material (LF models). With a wide variety of optional equipment, these models can be optimally adapted to your processes.

  • Tmax 1200 °C, 1300 °C, or 1400 °C
  • Five-sided heating for very good temperature uniformity
  • Heating elements on support tubes ensure free heat radiation and a long service life
  • Protection of bottom heating and flat stacking surface provided by embedded SiC plate in the floor
  • LH models: multilayered, fibre-free insulation of light refractory bricks and special backup insulation
  • LF models: high-quality fibre insulation with corner bricks for shorter heating and cooling times
  • Door with brick-on-brick seal, hand fitted
  • Short heating times due to high installed power
  • Side vent with bypass connection for exhaust pipe
  • Self-supporting arch for high stability and greatest possible protection against dust
  • Quick lock on door
  • Freely adjustable air slide intake in furnace floor
  • Stand included

Additional equipment

  • Parallel swinging door, pivots away from operator, for opening when hot
  • Lift door with electro-mechanic linear drive
  • Separate wall-mounting or floor standing cabinet for switchgear
  • Automatic vent flap
  • Cooling fan for shorter cycle times
  • Protective gas connector, sealed housing
  • Retort made of quartz glass for very clean atmosphere, quartz glass covered door with lid function
  • Manual or automatic gas supply system
  • Scale to measure weight reduction during annealing
ModelTmaxInner dimensions in mmVolumeOuter dimensions in mmPowerElectricalWeight
 °Cwdhin LWDHkWconnection*in kg
LH 15/1212002502502501557079011705.03-phase¹150
LH 30/1212003203203203064086012407.03-phase¹170
LH 60/12120040040040060720101013208.03-phase260
LH 120/1212005005005001208201110142012.03-phase340
            
LH 15/1313002502502501557079011707.03-phase¹150
LH 30/1313003203203203064086012408.03-phase¹170
LH 60/131300400400400607201010132011.03-phase260
LH 120/1313005005005001208201110142015.03-phase340
            
LH 15/1414002502502501557079011708.03-phase¹150
LH 30/14140032032032030640860124010.03-phase¹170
LH 60/141400400400400607201010132012.03-phase260
LH 120/1414005005005001208201110142018.03-phase340
            
LF 15/1313002502502501557079011707.03-phase¹130
LF 30/1313003203203203064086012408.03-phase¹150
LF 60/131300400400400607201010132011.03-phase230
LF 120/1313005005005001208201110142015.03-phase300
            
LF 15/1414002502502501557079011708.03-phase¹130
LF 30/14140032032032030640860124010.03-phase¹150
LF 60/141400400400400607201010132012.03-phase230
LF 120/1414005005005001208201110142018.03-phase300
¹Heating only between two phases *Please see page 70 for information on mains voltage

advancedmaterials_englisch catalogue Advanced Materials
(URL: http://www.nabertherm.nl/produkte/advancedmaterials/advancedmaterials_englisch.pdf)