High-Temperature Furnaces HTC 01/14 - HTCT 08/16 with SiC Rod Heating, optionally with Flap Door or Lift Door for Zirconia Sintering
 
 HTCT 08/15 with lift door HTC 08/14 with flap door and gas supply system

Furnace chamber with high-quality fibre materials and SiC heating rods on both sides of the furnace

Saggars with top lid

Spacers

HTC 01/14 - HTCT 08/16

Designed as table-top models with silicon carbide heating rods, these furnaces offer a variety of advantages when sintering Zirconia units. The large heating chamber, fast heat-up ramps and the apparent lack of chemical interactivity between zirconia and silicon carbide heating elements make these models the right choice for CAD/CAM manufacturing of zirconia restorations. The controller can be programmed for the individual zirconia material. The furnace is available in three different sizes. The 1 liter model can even be connected to the single phase power supply.

  • Tmax 1400 °C, 1500 °C, or 1600 °C
  • High-quality fibre materials, selected for the working temperature
  • Housing made of sheets of textured stainless steel
  • Double-walled housing for low external temperatures and high stability
  • Typical cycle time for zirconia firing at 1500 °C approx. 8 hours including cooling down
  • Optional flap door (HTC) which can be used as work platform or lift door (HTCT) with hot surface facing away from the operator
  • Adjustable air intake opening in the furnace door, exhaust air opening in the back wall
  • Switching system with solid-state-relays, power tuned to the SiC rods
  • Easy replacement of heating rods

Additional Equipment

  • Over-temperature limit controller with manual reset for thermal protection class 2 in accordance with EN 60519-2 as temperature limiter to protect the oven and load
  • Square saggar for charging of up to three layers
  • Spacer recommended to be placed under the bottom saggar for better temperature uniformity
  • Manual or automatic gas supply system
ModelTmaxInner dimensions in mmVolumeOuter dimensions in mmPowerElectricalWeightMax. number
 °Cwdhin LWDkWconnection*in kgof crowns
HTC, HTCT 01/1414009015015014707007502,2single-phase5540
HTC, HTCT 03/14140012021012034005355309,03-phase¹3070
HTC, HTCT 08/141400170290170845062057013,03-phase40120
             
HTC, HTCT 01/1515009015015014707007502,2single-phase5540
HTC, HTCT 03/15150012021012034005355309,03-phase¹3070
HTC, HTCT 08/151500170290170845062057013,03-phase40120
             
HTC, HTCT 01/1616009015015014707007502,2single-phase5540
HTC, HTCT 03/16160012021012034005355309,03-phase¹3070
HTC, HTCT 08/161600170290170845062057013,03-phase40120
¹Heating only between two phases *Please see page 14 for more information about mains voltage
²Plus maximum 270 mm for models HTCT when open